Shielding calculations for neutron calibration bunker using Monte Carlo code MCNP-4C

  • 1. Atomic Energy Commission, Damascus (Syrian Arab Republic), Division of Nuclear and Radiation Regulatory

Description

In this work, the dose arising from an Am-Be source of 108 neutron/sec strength located inside the newly constructed neutron calibration bunker in the National Radiation Metrology Laboratories, was calculated using MCNP-4C code. It was found that the shielding of the neutron calibration bunker is sufficient. As the calculated dose is not expected to exceed in inhabited areas 0.183 μSv/hr, which is 10 times smaller than the regulatory dose constraints. Hence, it can be concluded that the calibration bunker can house - from the external exposure point of view - an Am-Be neutron source of 109 neutron/sec strength. It turned out that the neutron dose from the source is few times greater than the photon dose. The sky shine was found to contribute significantly to the total dose. This contribution was estimated to be 60% of the neutron dose and 10% of the photon dose. The systematic uncertainties due to various factors have been assessed and was found to be between 4 and 10% due to concrete density variations; 15% due to the dose estimation method; 4 -10% due to weather variations (temperature and moisture). The calculated dose was highly sensitive to the changes in source spectra. The uncertainty due to the use of two different neutron spectra is about 70%.(author)

Availability note (English)

Available from INIS in electronic form

Files

39110487.pdf

Files (481.9 kB)

Name Size Download all
md5:2491b729a7fa8900645c8d1d50155ba1
481.9 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
28 p.
Report number
AECS-TN/RSS--763

Optional Information

Notes
12 refs., 7 tabs., 14 figs.