A water calorimeter for neutron dosimetry
Description
A water calorimeter has been used for the direct measurement of absorbed dose to water in a d(15)+Be neutron beam. The absorbed dose measured with the calorimeter was compared with that measured with an Exradin ionisation chamber, constructed of A-150 plastic. The doses measured by the ionisation chamber were calculated according to the European (ECNEU) protocol. Absorbed dose to tissue measured with the calorimeter was 4.3% lower than that measured with the ionisation chamber. Relative to ionisation chamber dosimetry, dose measurements with the calorimeter in the neutron beam were 9% lower than similar measurements in 4 and 9 MV photon beams. The significance of the results is discussed in terms of the heat defect in water. (author)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Volume
- 31
- Journal Issue
- 4
- Series
- Phys. Med. Biol.
- Journal Page Range
- 397-406
- ISSN
- 0031-9155
- CODEN
- PHMBA
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 17065654
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- CALORIMETRIC DOSEMETERS; COMPARATIVE EVALUATIONS; ERRORS; IONIZATION CHAMBERS; MEV RANGE 01-10; NEUTRON DOSIMETRY; PHOTON BEAMS; RADIOTHERAPY; WATER
- Descriptors DEC
- BEAMS; DOSEMETERS; DOSIMETRY; ENERGY RANGE; EVALUATION; HYDROGEN COMPOUNDS; MEASURING INSTRUMENTS; MEDICINE; MEV RANGE; OXYGEN COMPOUNDS; POLAR SOLVENTS; RADIATION DETECTORS; SOLVENTS; THERAPY