Published April 2007 | Version v1
Journal article

A patient-specific quality assurance program for the dose delivery of intensity modulated radiotherapy

  • 1. Cancer Center, Union Hospital, Tongji Medical College, Huazhong Univ. of Science and Technology, Wuhan (China)

Description

Objective: To establish a standard program of the dose delivery for every individual patient and to study the criterion of quality assurance in intensity modulated radiotherapy (IMRT). Methods: 33 nasopharynx cancer (NPC) IMRT treatment plan were designed, then transferred to the hybrid plan in a 3D treatment plan system (CMS Xio). The reference data were obtained. A ion chamber and Kodak EDR2 films were used to measure the absolute dose and the relative dose distribution, and difference between the reference value and measure value were evaluated with an advance criterion. Results: For 32 cases, the absolute dose value at the isocenter was below the tolerance limit of 5 %; for one case, the dose value exceeded the range, but re-meet the demand through adjusting the measure point to a flat region of dose distribution. For relative dose distribution, all cases meet the tolerance limit of 5 % and 3 mm distance difference. Conclusions: To establish a standard program for individual patient is very important for the exact implement of IMRT plan. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiological Medicine and Protection
Journal Volume
27
Journal Issue
2
Journal Page Range
p. 170-172
ISSN
0254-5098

Optional Information

Notes
1 fig., 10 refs.