Published February 1981
| Version v1
Journal article
Comments on 'Comparison of the Bragg-Gray theory corollaries to fast-neutron cavity ionization measurements at 14.8 MeV' by F.H. Attix et al.; and reply
Creators
- 1. Illinois Univ., Urbana (USA). Dept. of Nuclear Engineering
- 2. Wisconsin Univ., Madison (USA). Dept. of Radiology
Description
The author examines work by Attix et al on the assumption that in ionization detectors used for neutron dosimetry. Bragg-Grey Theory requirements are satisfied for 14.8 MeV neutrons. The conclusion is that Bragg-Grey Theory can only be extended to high energy neutron irradiation of inhomogeneous detectors if the gas cavity size or gas pressure are reduced such that the domestic ionization is an insignificant part of the total. Attix replies. (U.K.)
Additional details
Publishing Information
- Journal Title
- Health Phys.
- Journal Volume
- 40
- Journal Issue
- 2
- Series
- Health Phys.
- Journal Page Range
- 259-263
- ISSN
- 0017-9078
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United Kingdom
- INIS RN
- 12619056
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BRAGG GRAY CHAMBERS; FAST NEUTRONS; IONIZATION; MEV RANGE 10-100; NEUTRON DOSIMETRY
- Descriptors DEC
- BARYONS; DOSEMETERS; DOSIMETRY; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; HADRONS; IONIZATION CHAMBERS; MEASURING INSTRUMENTS; MEV RANGE; NEUTRONS; NUCLEONS; RADIATION DETECTORS