Published September 1982 | Version v1
Journal article

On a mosaic graphite spectrometer without collimators

Description

An X-ray spectrometer was constructed for the detection of very weak lines having a large energy shift from the primary (Rayleigh) line. The spectrometer has no collimators and uses a graphite mosaic crystal as an analyser. The dependence of the resolution and intensity on one another is examined and both are calculated as a function of the scatterer-analyzer distance. The spectrometer was tested in X-ray inelastic scattering experiments

Additional details

Publishing Information

Journal Title
Can. J. Phys.
Journal Volume
60
Journal Issue
9
Series
Can. J. Phys.
Journal Page Range
1241-1246
ISSN
0008-4204

INIS

Country of Publication
Canada
Country of Input or Organization
Canada
INIS RN
14762123
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CRYSTALS; GRAPHITE; RESOLUTION; X-RAY SPECTROMETERS
Descriptors DEC
CARBON; ELEMENTS; MEASURING INSTRUMENTS; NONMETALS; SPECTROMETERS