Published November 2011 | Version v1
Journal article

Comparison of PET performance in clinical examination among PET facilities

  • 1. Iwate Medical Univ., Cyclotron Research Center, Takizawa, Iwate (Japan)
  • 2. Tohoku Univ., School of Engineering, Sendai, Miyagi (Japan)

Description

There are several guidelines for positron emission tomography (PET) performance evaluation. Most of them require a lot of time and strict conditioning. Here we selected 5 items from the guidelines with some adjustments on which we evaluate and compare PET performances of seven facilities for the purpose of clarifying why and how PET quantitative values are different among facilities. The results show that the data vary even with the same type of PET machine when the image reconstruction method is different. This suggests that the terms of data reconstruction have to be considered as well as the difference of PET performance when we refer PET data of other facilities for clinical purpose. (author)

Availability note (English)

Available from http://dx.doi.org/10.3769/radioisotopes.60.473

Additional details

Identifiers

Publishing Information

Journal Title
Radioisotopes (Tokyo)
Journal Volume
60
Journal Issue
11
Journal Page Range
p. 473-486
ISSN
0033-8303