Published October 1987
| Version v1
Journal article
A Josephson array voltage standard at 10 V
- 1. Electromagnetic Technology Div., National Bureau of Standards, Boulder, CO (USA)
Description
The technology of Josephson voltage standards has been extended to an array of 14 184 junctions which is capable of generating over 150 000 quantized voltage levels spanning the range from -12 to +12V. This makes possible the direct calibration of 10-V Zener reference standards without the use of a voltage divider
Additional details
Publishing Information
- Journal Title
- IEEE Electron Device Lett.
- Journal Volume
- EDL-8
- Journal Issue
- 10
- Series
- IEEE Electron Device Lett.
- Journal Page Range
- 449-450
- ISSN
- 0741-3106
- CODEN
- EDLED
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 19086367
- Subject category
- S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
- Descriptors DEI
- CALIBRATION; ELECTRIC POTENTIAL; JOSEPHSON JUNCTIONS; JUNCTION DIODES; PERFORMANCE; SPECIFICATIONS; VOLTAGE REGULATORS
- Descriptors DEC
- SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; SEMICONDUCTOR JUNCTIONS