Published 2002 | Version v1
Journal article

LiF:Mg,Cu,P 'pin worms': miniature detectors for brachytherapy dosimetry

Description

Dose measurements in brachytherapy 192Ir implants are often difficult due to large dose gradients and complex photon spectra. Therefore, tissue-equivalent detectors with a high spatial resolution, such as the highly promising LiF:Mg,Cu,P thermoluminescent detectors (TLDs) are required. It was the aim of the present work to ascertain if miniature LiF:Mg,Cu,P TLDs can effectively measure the dose distribution around 192Ir implants. 'Pin worm' TLDs (type MCP, diameter 0.6 mm, length 2 mm) were GR-200 R (SSDL, Beijing) rods cut in half. The TLDs were tested for reproducibility and energy dependence using high dose rate and low dose rate brachytherapy units. 192Ir measurements were performed in a tissue equivalent phantom accommodating hollow needles and catheters routinely used in brachytherapy. Pin worms had a reproducibility of less than ±2% (1 SD) and a detection limit of less than 10 μGy. The small dimensions of the pin worms allowed their placement within brachytherapy needles and catheters. The measured relative dose distribution was in good agreement with the predictions of a computerised treatment planning system (ADAC Pinnacle); however, limitations in the TLD energy correction did not allow for absolute dose comparison. (author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
101
Journal Issue
1-4
Journal Page Range
p. 407-410
ISSN
0144-8420
CODEN
RPDODE

Conference

Title
13. international conference on solid state dosimetry
Dates
9-13 Jul 2001
Place
Athens (Greece)