Dosimetric characteristics of the Best registered double-wall 103Pd brachytherapy source

  • 1. Department of Radiation Medicine, University of Kentucky Medical Center, Lexington, Kentucky 40536-0084 (United States)

Description

103Pd and 125I brachytherapy sources are being used for interstitial implants in tumor sites such as the prostate. Recently, a double-wall 103Pd source has been introduced, which has a design different from that of sources presently on the market. Dosimetric characteristics (dose rate constant, radial dose function, and anisotropy function) of this source were experimentally and theoretically determined following the AAPM Task Group 43 recommendations and were related to the October 10, 2000 revision of the NIST 1999 SK Standard for 103Pd. Measurements were performed in a Solid Water trade mark sign phantom using LiF thermoluminescent dosimeters. For these measurements, slabs of Solid Water trade mark sign phantom material were machined to accommodate the source and LiF TLD chips of dimensions (3.1x3.1x0.8 mm3) and (1.0x1.0x1.0 mm3). The TLD chips were surrounded by at least 10 cm of Solid Water trade mark sign phantom material to provide full scattering conditions. The Monte Carlo simulations were performed in Solid Water trade mark sign and liquid water using the PTRAN code. The results of this investigation show an excellent agreement (within 5%) between the measured (0.67±8% cGy h-1 U-1) and calculated (to be 0.65±3% cGy h-1 U-1) dose rate constant in Solid Water trade mark sign . The Monte Carlo calculated dose rate constant of the Best registered 103Pd in water was found to be 0.67±0.02 cGy h-1 U-1. The radial dose function, g(r), of the new 103Pd source was measured at distances ranging from 0.5 and 7 cm using LiF TLD in Solid Water trade mark sign phantom material. Moreover, the radial dose function of the new source was calculated in liquid water and Solid Water trade mark sign at distances ranging from 0.1 to 7 cm using the PTRAN Monte Carlo Code. The anisotropy function, F(r,θ), of the new 103Pd source was also measured in Solid Water trade mark sign and calculated in both Solid Water trade mark sign and water phantom material. From the anisotropy functions, the anisotropy factors, and anisotropy constant were calculated for each medium. The results indicated that the measured anisotropy constant of the Best registered 103Pd source in Solid Water trade mark sign was 0.89±5%. Complete dosimetric data are described in this manuscript

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
28
Journal Issue
12
Journal Page Range
p. 2568-2575
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2001 American Association of Physicists in Medicine.