Published February 22, 2003 | Version v1
Journal article

New triiodomercurate-modified carbon paste electrode for the potentiometric determination of mercury

Description

A new tetrazolium-triiodomercurate-modified carbon paste electrode has been described for the sensitive and selective determination of mercury. The electrode shows a stable, near-Nernstian response for 1x10-3 to 6x10-6 M [HgI3]- at 25 deg. C over the pH range of 4.0-9.0, with an anionic slope of 55.5±0.4 mV. The lower detection limit is 4x10-6 M with a fast response time of 30-50 s. Selectivity coefficients of a number of interfering anions and iodo complexes of some metal ions have been estimated. The interference from many of the investigated ions is negligible. The determination of 1-200 μg/ml of mercury in aqueous solutions shows an average recovery of 98.5% and a mean relative standard deviation of 1.6% at 50.0 μg/ml. The direct determination of mercury in spiked wastewater, metal amalgams and dental alloy gave results that compare favorably with those obtained by the cold vapor atomic absorption spectrometric method. Potentiometric titration of mercury and phenylmercury acetate with standard potassium iodide has been monitored using the developed triiodomercurate-carbon paste electrode (CPE) as an end point indicator electrode

Additional details

Identifiers

DOI
10.1016/S0003-2670(02)01520-9;
arXiv
arXiv:hep-ph/0112170v1;
PII
S0003267002015209;

Publishing Information

Journal Title
Analytica Chimica Acta
Journal Volume
478
Journal Issue
2
Journal Page Range
p. 329-335
ISSN
0003-2670
CODEN
ACACAM

Optional Information

Copyright
Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.