Purpose
To compare, in an interventional radiology setting, peak skin doses (PSDs) delivered as calculated using a dedicated software tool and as measured using radiochromic film. To assess the utility of this dose calculation software tool in routine clinical practice.Materials and Methods
First, radiochromic films were positioned on the examination table in the back of an adult anthropomorphic phantom to measure PSD, and X-ray examinations were simulated. Then, films were again positioned in the patient's back for 59 thoracic or abdominopelvic endovascular interventions. The results obtained with the radiochromic films were taken as a reference and were statistically compared with those of the software.Results
With measured PSDs ranging from 100 to 7000 mGy, the median software–film difference was 8.5%. Lin's concordance coefficient was 0.98 [0.97; 0.99] (p < 0.001), meaning that concordance was excellent between the two methods. For the films where PSD exceeded 1000 mGy, the median difference in the measured value was 8.7% [− 1.3; 21.1], with a maximum discrepancy of 34%. Lin's concordance coefficient was 0.98 [0.96; 1] (p < 0.001), meaning that concordance was excellent between the two methods.Conclusion
Comparison between radiochromic films and the software tool showed that the software is a suitable tool for a simple and reliable estimation of PSD. The software seems to be a good alternative to films, whose use remains complex.Additional details
Identifiers
Publishing Information
- Journal Title
- Cardiovascular and Interventional Radiology
- Journal Volume
- 41
- Journal Issue
- 5
- Journal Page Range
- p. 762-771
- ISSN
- 0174-1551
- CODEN
- CAIRDG
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 49101017
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; COMPUTER CODES; FILM DOSIMETRY; PATIENTS; SKIN; X RADIATION
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DOSIMETRY; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIATIONS; RADIOLOGY