Description

Purpose

To compare, in an interventional radiology setting, peak skin doses (PSDs) delivered as calculated using a dedicated software tool and as measured using radiochromic film. To assess the utility of this dose calculation software tool in routine clinical practice.

Materials and Methods

First, radiochromic films were positioned on the examination table in the back of an adult anthropomorphic phantom to measure PSD, and X-ray examinations were simulated. Then, films were again positioned in the patient's back for 59 thoracic or abdominopelvic endovascular interventions. The results obtained with the radiochromic films were taken as a reference and were statistically compared with those of the software.

Results

With measured PSDs ranging from 100 to 7000 mGy, the median software–film difference was 8.5%. Lin's concordance coefficient was 0.98 [0.97; 0.99] (p < 0.001), meaning that concordance was excellent between the two methods. For the films where PSD exceeded 1000 mGy, the median difference in the measured value was 8.7% [− 1.3; 21.1], with a maximum discrepancy of 34%. Lin's concordance coefficient was 0.98 [0.96; 1] (p < 0.001), meaning that concordance was excellent between the two methods.

Conclusion

Comparison between radiochromic films and the software tool showed that the software is a suitable tool for a simple and reliable estimation of PSD. The software seems to be a good alternative to films, whose use remains complex.

Additional details

Identifiers

Publishing Information

Journal Title
Cardiovascular and Interventional Radiology
Journal Volume
41
Journal Issue
5
Journal Page Range
p. 762-771
ISSN
0174-1551
CODEN
CAIRDG

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
49101017
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; COMPUTER CODES; FILM DOSIMETRY; PATIENTS; SKIN; X RADIATION
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DOSIMETRY; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIATIONS; RADIOLOGY

Optional Information

Copyright
Copyright (c) 2018 Springer Science+Business Media, LLC, part of Springer Nature and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE)
Notes
http://www.springer-ny.com