Validation of a new grid-based Boltzmann equation solver for dose calculation in radiotherapy with photon beams

  • 1. Department of Radiation Physics, MD Anderson Cancer Center, Houston, TX 77030 (United States)
  • 2. Transpire, Inc., Gig Harbor, WA 98335 (United States)

Description

A new grid-based Boltzmann equation solver, Acuros(TM), was developed specifically for performing accurate and rapid radiotherapy dose calculations. In this study we benchmarked its performance against Monte Carlo for 6 and 18 MV photon beams in heterogeneous media. Acuros solves the coupled Boltzmann transport equations for neutral and charged particles on a locally adaptive Cartesian grid. The Acuros solver is an optimized rewrite of the general purpose Attila software, and for comparable accuracy levels, it is roughly an order of magnitude faster than Attila. Comparisons were made between Monte Carlo (EGSnrc) and Acuros for 6 and 18 MV photon beams impinging on a slab phantom comprising tissue, bone and lung materials. To provide an accurate reference solution, Monte Carlo simulations were run to a tight statistical uncertainty (σ ∼ 0.1%) and fine resolution (1-2 mm). Acuros results were output on a 2 mm cubic voxel grid encompassing the entire phantom. Comparisons were also made for a breast treatment plan on an anthropomorphic phantom. For the slab phantom in regions where the dose exceeded 10% of the maximum dose, agreement between Acuros and Monte Carlo was within 2% of the local dose or 1 mm distance to agreement. For the breast case, agreement was within 2% of local dose or 2 mm distance to agreement in 99.9% of voxels where the dose exceeded 10% of the prescription dose. Elsewhere, in low dose regions, agreement for all cases was within 1% of the maximum dose. Since all Acuros calculations required less than 5 min on a dual-core two-processor workstation, it is efficient enough for routine clinical use. Additionally, since Acuros calculation times are only weakly dependent on the number of beams, Acuros may ideally be suited to arc therapies, where current clinical algorithms may incur long calculation times.

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/55/3/002

Additional details

Identifiers

DOI
10.1088/0031-9155/55/3/002;
PII
S0031-9155(10)26042-6;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
55
Journal Issue
3
Journal Page Range
p. 581-598
ISSN
0031-9155
CODEN
PHMBA7