Performance evaluation of a computed radiography imaging device using a typical 'front side' and novel 'dual side' readout storage phosphors

  • 1. Diagnostic Radiology Department, Mayo Clinic, Rochester, Minnesota 55905 (United States)

Description

The Fourier-space modulation transfer function (MTF), normalized noise power spectrum (nNPS), and detective quantum efficiency (DQE) of a computed radiography (CR) x-ray imaging device were measured. Two different storage phosphor screens were used in conjunction with a single CR reader (Fuji, Clearview, CS). One of the storage phosphor plates (ST-BD) had a clear backing material which allowed 'dual side read' of the latent image from both the 'front' and 'back' sides of the phosphor. The other phosphor plate had a light occluding backing material, limiting the readout to front side only (ST-VI). The standard RQA-5 beam quality was used. The MTF was measured using a 1 mm thick tungsten edge device. Compared to the ST-55BD phosphor, the ST-VI phosphor was found to have modestly higher MTF at all spatial frequencies. The nNPS(f) and DQE(f) were measured for nominal incident exposure levels ranging from 0.1 to 10 mR. The dual side read phosphor demonstrated superior DQE, especially at low spatial frequencies. At the frequency 0.5 cycles/mm, the DQE values for the 1 mR exposure were 0.36 and 0.21 for the ST-55BD and ST-VI phosphor plates, respectively. The differences between the spatial-frequency dependent DQE of the two plates can be attributed to the increased signal collection efficiency of the dual side read plates and differences in storage phosphor structure noise

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
33
Journal Issue
2
Journal Page Range
p. 290-296
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2006 American Association of Physicists in Medicine