Published May 5, 1976 | Version v1
Report Open

The rapid determination by x-ray-fluorescence spectrometry of tungsten in tungsten ores and concentrates

Description

A method is presented that is applicable to tungsten concentrations above 0,1 per cent. Matrix variations are suppressed by a tenfold to fiftyfold dilution of the sample material during fusion, which eliminates any mineralogical effects and produces a clear glass disc that is used direct for intensity measurements. Residual matrix variations are compensated for by application of the background-ratio technique, and the intensity measurements are evaluated by comparison with synthetic standards, the calculation being done on-line by use of an Olivetti Programma 602 desk-top computer. An annotated computer programme is provided for calculation of the curve of the calibration graph and evaluation of the sample intensities. The method is rapid and has a coefficient of variation of 1,5 per cent at a level of 50 per cent tungsten when used under routine conditions. Comparison with the results of a wet-chemical method showed the proposed method to have no bias

Availability note (English)

MF available from INIS under the Report Number.

Files

16001461.pdf

Files (212.2 kB)

Name Size Download all
md5:913adb15ba0a175cb93300de95da520d
212.2 kB Preview Download

Additional details

Additional titles

Subtitle (English)
Laboratory method no. 74/4

Publishing Information

ISBN
0 86999 277 5
Imprint Pagination
12 p.
Report number
NIM--1817