The usefulness of large sample size patient dose audits for optimisation of CT automatic exposure control (AEC) settings
- 1. Medical Physics Service, Queen's Centre, Castle Hill Hospital, Hull University Teaching Hospitals NHS Trust, Castle Road, Hull, HU16 5JQ (United Kingdom)
Description
The aim of this study was to demonstrate the usefulness of large sample size patient dose audits for optimisation of CT automatic exposure control (AEC) settings, even when the investigation is limited to only three scanners at a single institution. Pre-optimisation patient dose audits of common CT examinations (n > 200 for each protocol) on three CT scanners (two Philips Brilliance and one Toshiba Aquilion) using radiology information system (RIS) data were conducted showing sub-optimal CT AEC performance on the Toshiba scanner. Based on these results, an optimisation exercise was carried out on the non-optimally performing scanner by phantom measurement and investigation of system configuration. Post-optimisation patient dose audits were subsequently carried out to assess the success of the optimisation exercise demonstrating standardisation of doses; median dose-length-product values were reduced by up to 43% on the sub-optimal scanner without any adverse effect on clinical image quality. This study has demonstrated that large sample patient dose audits using RIS data can be instrumental in identifying and rectifying sub-optimal CT AEC performance, even when the investigation is limited to only three scanners at a single institution. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6498/ab3894Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Radiological Protection
- Journal Volume
- 39
- Journal Issue
- 3
- Journal Page Range
- p. 938-949
- ISSN
- 0952-4746
- CODEN
- JRPREA
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 51078295
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; EXERCISE; IMAGES; OPTIMIZATION; PATIENTS; PERFORMANCE; PHANTOMS; RADIATION DOSES; RADIOLOGY; STANDARDIZATION
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; DOSES; MEDICINE; MOCKUP; NUCLEAR MEDICINE; STRUCTURAL MODELS; TOMOGRAPHY