A new method for matrix corrections by radioisotope excited X-ray fluorescence
Description
A method has been investigated which allows calculations of the absorption correction factors in the energy range 1.0 to 5.0 keV using 55Fe excitation source. From the experimentally-measured absorption factors between 6.0 and 20.0 keV using 109Cd excitation source, the coefficients at lower energies with the associated absorption jumps are generated through an iterative procedure. (author)
Additional details
Publishing Information
- Journal Title
- X-Ray Spectrom., XRS
- Journal Volume
- 12
- Journal Issue
- 4
- Series
- X-Ray Spectrom., XRS.
- Journal Page Range
- 148-149
- ISSN
- 0049-8246
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 15005174
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- ABSORPTION; CADMIUM 109; CORRECTIONS; EXCITATION; IRON 55; ITERATIVE METHODS; KEV RANGE 01-10; KEV RANGE 10-100; RADIATION SOURCES; X-RAY FLUORESCENCE ANALYSIS
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; CADMIUM ISOTOPES; CHEMICAL ANALYSIS; ELECTRON CAPTURE RADIOISOTOPES; ENERGY RANGE; ENERGY-LEVEL TRANSITIONS; EVEN-ODD NUCLEI; INTERMEDIATE MASS NUCLEI; IRON ISOTOPES; ISOTOPES; KEV RANGE; NONDESTRUCTIVE ANALYSIS; NUCLEI; RADIOISOTOPES; X-RAY EMISSION ANALYSIS; YEARS LIVING RADIOISOTOPES