Published January 1, 2020 | Version v1
Journal article

Reversed-slit spectroscopy method for in situ measurement of H isotopes on plasma facing material

  • 1. Forschungszentrum Jülich GmbH, Institut für Energie- und Klimaforschung - Plasmaphysik, 52425 Jülich (Germany)

Description

Optic spectroscopy methods are employed for plasma facing component (PFC) material diagnostics and H isotopes retention on various platforms. A common issue of those spectroscopy methods is the measurement sensitivity. Various improvement methods have been investigated. In the current work, a 'reversed-slit' method is explored, aiming to maximise signal input while retain a reasonably good line resolution. A thin metal wire with a diameter that is comparable to a normal slit width is placed in front of an open slit. By opening the slit the input of the emission light is maximised. The thin wire casts a line of shadow on the spectral image. The shadow line is used as the 'characteristic line' to identify and separate the spectral line of each H isotope. The result shows that an open slit of 3 mm width can bring an intensity increase of 20–30 times compared to a 100 μm wide normal slit without compromising the resolution.

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/15/01/C01007

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
15
Journal Issue
01
Journal Page Range
p. C01007
ISSN
1748-0221

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52088912
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CURRENTS; FIRST WALL; HYDROGEN ISOTOPES; METALS; RESOLUTION; SENSITIVITY; SIGNALS; SPECTROSCOPY; WIDTH; WIRES
Descriptors DEC
DIMENSIONS; ELEMENTS; ISOTOPES; THERMONUCLEAR REACTOR WALLS