Published 1980 | Version v1
Journal article

Balloon catheters

  • 1. Soedersjukhuset, Stockholm (Sweden)

Description

Balloon catheters with latex balloons were inflated in plastic tubes of different diameters. The volume at rupture of the balloons was measured. The pressure inside the balloon and that caused by the balloon against the tube walls were recorded. Measures to avoid high pressure on the vessel wall, and balloon rupture, are discussed. (Auth.)

Additional details

Additional titles

Subtitle (English)
In vitro experiments on pressure influence and risk of rupture

Publishing Information

Journal Title
Acta Radiol., Diagn.
Journal Volume
21
Journal Issue
1
Series
Acta Radiol., Diagn.
Journal Page Range
47-52
ISSN
0567-8056

INIS

Country of Publication
Sweden
Country of Input or Organization
Sweden
INIS RN
11552511
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; BLOOD FLOW; BLOOD VESSELS; DIAGNOSIS; IN VITRO
Descriptors DEC
BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; MEDICINE; ORGANS