Published December 10, 1990 | Version v1
Journal article

Suboptimal filtering of 1/f-noise in detector charge measurements

  • 1. Istituto Nazionale di Fisica Nucleare, Milan (Italy)
  • 2. Politecnico di Milano (Italy). Dipartimento di Elettronica
  • 3. Pavia Univ. (Italy). Dipt. di Elettronica

Description

Processing of 1/f-noise by suboptimal filters in detector charge measurements is considered. The achievable values of equivalent noise charge are compared to those provided by the optimum filter. Some aspects of filters processing 1/f-noise, like the degradation in the accuracy of charge measurments due to a bipolar or to a unipolar, flat-topped weighting function are described. Several conclusions are expected to be useful in the design of analog processors for radiation detectors when the front-end 1/f-noise is not negligible. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research, Section A
Journal Volume
297
Journal Issue
3
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
467-478
ISSN
0168-9002
CODEN
NIMAE