Dosimetric study of chest computerized tomography in childhood using phantoms

  • 1. Universidade Federal de Minas Gerais (DEN/UFMG), Belo Horizonte, MG (Brazil). Dept. de Engenharia Nuclear

Description

The dissemination of computed tomography (CT) promoted a significant increase in the dose absorbed by patients due to the diagnosis. Therefore, it is essential to improve protocols, seeking lower doses, without prejudice to the diagnostic quality of the image. The risks of stochastic effects are higher for children due to tissue radiosensitivity combined with longer life expectancy. In this work, a cylindrical simulator object representing an adult thorax made of polymethyl methacrylate was used, and simulated objects of the same material were developed, in oblong shape including axillary regions, based on the dimensions of the chest of a pediatric patient. A comparative study was performed between chest scans performed on two CT scans at different radiology services. The central slice of the simulating objects was irradiated successively and, using a pencil-type ionization chamber, absorbed dose measurements were made at five distinct points of each simulating object. From the measurements, the values of weighted and volumetric dose indices were obtained for the 10 cm scan of the central region of the objects, in helical mode. The scans were performed with the chest acquisition protocols used by the radiodiagnostic services, both for a 120 kV X-ray tube supply voltage. This study compared the absorbed dose variation between patients with different chest volumes and the patient dose variation between two devices when used for imaging with the same diagnostic objective, with the purpose of proposing optimized protocols. The results indicated that the protocols defined for children are oversized and need to be corrected to optimize the dose deposited in these patients

Files

50061033.pdf

Files (581.8 kB)

Name Size Download all
md5:07cf581980f704a87c771c8417d664c4
581.8 kB Preview Download

Additional details

Additional titles

Original title (Portuguese)
Estudo dosimétrico de tomografia computadorizada de tórax na infância usando objetos simuladores

Publishing Information

Imprint Pagination
11 p.
Report number
INIS-BR--22701

Conference

Title
week on nuclear engineering and radiation sciences
Acronym
4. SENCIR 2018
Dates
6-8 Nov 2018
Place
Belo Horizonte, MG (Brazil)