Influence of the Integral Quality Monitor transmission detector on high energy photon beams. A multi-centre study
- 1. Institute of Oncology, Ljubljana (Slovenia). Dept. of Radiation Physics
- 2. Lake Constance Radiation Oncology Center, Singen and Friedrichshafen (Germany)
- 3. Wuerzburg Univ. (Germany). Dept. of Radiation Oncology
Description
The influence of the Integral Quality Monitor (IQM) transmission detector on photon beam properties was evaluated in a preclinical phase, using data from nine participating centres: (i) the change of beam quality (beam hardening), (ii) the influence on surface dose, and (iii) the attenuation of the IQM detector. For 6 different nominal photon energies (4 standard, 2 FFF) and square field sizes from 1 x 1 cm2 to 20 x 20 cm2, the effect of IQM on beam quality was assessed from the PDD20,10 values obtained from the percentage dose depth (PDD) curves, measured with and without IQM in the beam path. The change in surface dose with/without IQM was assessed for all available energies and field sizes from 4 x 4 cm2 to 20 x 20 cm2. The transmission factor was calculated by means of measured absorbed dose at 10 cm depth for all available energies and field sizes. (i) A small (0.11-0.53%) yet statistically significant beam hardening effect was observed, depending on photon beam energy. (ii) The increase in surface dose correlated with field size (p < 0.01) for all photon energies except for 18 MV. The change in surface dose was smaller than 3.3% in all cases except for the 20 x 20 cm2 field and 10 MV FFF beam, where it reached 8.1%. (iii) For standard beams, transmission of the IQM showed a weak dependence on the field size, and a pronounced dependence on the beam energy (0.9412 for 6 MV to 0.9578 for 18 MV and 0.9440 for 6 MV FFF; 0.9533 for 10 MV FFF). The effects of the IQM detector on photon beam properties were found to be small yet statistically significant. The magnitudes of changes which were found justify treating IQM either as tray factors within the treatment planning system (TPS) for a particular energy or alternatively as modified outputs for specific beam energy of linear accelerators, which eases the introduction of the IQM into clinical practice.
Additional details
Publishing Information
- Journal Title
- Zeitschrift fuer Medizinische Physik
- Journal Volume
- 27
- Journal Issue
- 3
- Journal Page Range
- p. 232-242
- ISSN
- 0939-3889
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 48083867
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CORRELATIONS; DEPTH DOSE DISTRIBUTIONS; DISPERSION RELATIONS; ENERGY DEPENDENCE; LINEAR ACCELERATORS; MONITORING; ON-LINE CONTROL SYSTEMS; PHOTON BEAMS; RADIATION DOSES; TRANSITION RADIATION DETECTORS; TRANSMISSION
- Descriptors DEC
- ACCELERATORS; BEAMS; CONTROL SYSTEMS; DOSES; MEASURING INSTRUMENTS; ON-LINE SYSTEMS; RADIATION DETECTORS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DOSE DISTRIBUTIONS