Published July 15, 2007 | Version v1
Journal article

Radiation Exposure During Head Repositioning With the Automatic Positioning System for Gamma Knife Radiosurgery

  • 1. Department of Therapeutic Radiology, University of Minnesota, Minneapolis, MN (United States)

Description

Purpose: To measure radiation exposure to a patient during head repositioning with the automatic positioning system (APS) for Gamma Knife radiosurgery. Methods and Materials: A 16-cm diameter spherical solid phantom, provided by the manufacturer, was mounted to the APS unit using a custom-made holder. A small-volume ionization chamber (0.07-cm3 volume) was placed at the center of the phantom. We recorded the temporal variation of ionization current during the entire treatment. Measurements were made for 3 test cases and 7 clinical cases. Results: The average transit time between successive shots, during which the APS unit was moving the phantom for repositioning the shot coordinates, was 20.5 s for 9 cases. The average dose rate, which was measured at the center of the phantom and at a point outside the shot location, was 0.36 ± 0.09 cGy/min when the beam output was approximately 3.03 Gy/min for the 18-mm collimator helmet. Hence, the additional intracranial radiation dose during the APS-driven head repositioning between two successive shots (or APS transit dose) was 0.12 ± 0.050 cGy. The APS transit dose was independent of the helmet size and the position of shots within the phantom relative to the measurement point. Conclusion: The head repositioning with the APS system adds a small but not negligible dose to the dose expected for the manual repositioning method

Additional details

Identifiers

DOI
10.1016/j.ijrobp.2007.03.052;
PII
S0360-3016(07)00568-8;

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
68
Journal Issue
4
Journal Page Range
p. 1207-1211
ISSN
0360-3016
CODEN
IOBPD3

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
38104852
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BEAMS; COLLIMATORS; DOSE RATES; HEAD; IONIZATION CHAMBERS; PHANTOMS; RADIATION DOSES; RADIOTHERAPY; SURGERY
Descriptors DEC
BODY; DOSES; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIOLOGY; STRUCTURAL MODELS; THERAPY

Optional Information

Copyright
Copyright (c) 2007 Elsevier Science B.V., Amsterdam, Netherlands, All rights reserved.