Deadtime measurement and associated statistical uncertainty by means of a two-detector paired-source method

  • 1. National Accelerator Centre, Faure (South Africa)

Description

A brief description of the two-detector paired-source method for determining the deadtime of a counting system is given. A very simple formula for accurately estimating the statistical uncertainty of the deadtime value is derived. The error formula has been carefully checked against experimental measurements. (author)

Additional details

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
42
Journal Issue
9
Series
Appl. Radiat. Isot.
Journal Page Range
811-814
ISSN
0883-2889
CODEN
ARISE

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
23002523
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
DATA COVARIANCES; DEAD TIME; ERRORS; RADIATION DETECTORS; RADIATION SOURCES
Descriptors DEC
MEASURING INSTRUMENTS; TIMING PROPERTIES