GAMMASPHERE: Elimination of ballistic deficit by using a quasi-trapezoidal pulse shaper

Description

Gammasphere uses an spherical array of very large (7.2cm dia.) germanium detectors and only high-multiplicity events are studied. To achieve a reasonable coincidence rate, the individual detector channels must handle high rates with minimum pile-up losses. Ten microseconds was chosen as the total processing time for a signal which means that the shaped signal peaks in about 4us. The combination of short pulse shaping and the fluctuating long charge collection times (up to 400ns) in the detectors exaggerates the energy resolution degradation due to ballistic deficit effects. We describe a method of producing a flat-topped pulse with a simple time-invariant network that satisfies GAMMASPHERE requirements and eliminates ballistic deficit effects

Availability note (English)

MF available from INIS under the Report Number; Also available from OSTI as DE94003252; NTIS; US Govt. Printing Office Dep.

Files

25029080.pdf

Files (173.0 kB)

Name Size Download all
md5:9f6ce9cd436def921de6d73abca21801
173.0 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
5 p.
Report number
LBL--33989

Conference

Title
Nuclear science symposium.
Dates
2-5 Nov 1993.
Place
San Francisco, CA (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
25029080
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
GAMMA DETECTION; GE SEMICONDUCTOR DETECTORS; PULSE SHAPERS; WAVE FORMS
Descriptors DEC
DETECTION; ELECTRONIC CIRCUITS; MEASURING INSTRUMENTS; PULSE CIRCUITS; RADIATION DETECTION; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS

Optional Information

Contract/Grant/Project number
Contract AC03-76SF00098
Funding organization
USDOE, Washington, DC (United States).
Secondary number(s)
CONF-931107--11.