Universal device for thermophysical experiments in a helium Dewar flask

Description

The construction of a device for thermophysical experiments and measurement of the thermal conductivity are described. The cryostat if 39mm in diameter, designed as an insert for a helium Dewar flask. The insert is equipped with quick-diassembling low-temperature vacuum seal, constructed as a lapped taper surface connection of a stainless steel parts. An express method for measuring the thermal conductivity using this insert is described. In the temperature interval from 5 K to 13K thermal conductivities of the stainless steel 12x18H10T, of electrical steel E-310, the epoxy adhesive EG-80, silicate glue with fillers are measured 2% absolute error for the method is estimated

Files

19102773.pdf

Files (208.2 kB)

Name Size Download all
md5:3c2a9476dd5c371e38f02f957b13c83f
208.2 kB Preview Download

Additional details

Additional titles

Original title (Russian)
Universal'noe ustrojstvo dlya teplofizicheskikh izmerenij v gelievom d'yuare

Publishing Information

Imprint Pagination
8 p.
Report number
JINR--8-87-708

Optional Information

Notes
10 refs.; 6 figs.; submitted to the journal Prib. Tekh. Ehksp.