Published November 1, 2018 | Version v1
Journal article

Experimental study of one step linear Gauss-Newton algorithm for improving the quality of image reconstruction in high-speed Electrical Impedance Tomography (EIT)

  • 1. Physics Department, Institut Teknologi Sepuluh Nopember Kampus ITS Sukolilo - Surabaya, 60111, East Jawa (Indonesia)

Description

One step linear Gauss-Newton was well known as one of image reconstruction algorithm in Electrical Impedance Tomography (EIT). In the present study, we proposed the mathematical model of one step linear Gauss-Newton for processing the experimental data for achieving the best image reconstruction. Two ways were conducted to solve this study. The first one, we build EIT hardware that mainly controlled by Single Board Computer (SBC) to produce a constant current at 1 mA 20 kHz that will be injected to the 16 electrodes of the practical phantom with speed of 1 ms/data for acquiring the data of voltage boundary from that practical phantom when using aluminium as anomaly. The second one, we processed the data of voltage boundary using one-step linear Gauss-Newton to achieve the best image reconstruction with modified the exponent (p) of Regularization matrix that varied at 0 − 1 by the increment of 0.1 and hyperparameter value (λ). The result showed that the best image reconstruction obtained at p = 0.7 and λ = 8 for this algorithm. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/1120/1/012067

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
1120
Journal Issue
1
Journal Page Range
[7 p.]
ISSN
1742-6596

Conference

Title
8. International Conference on Theoretical and Applied Physics
Dates
20-21 Sep 2018
Place
Medan (Indonesia)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
53019080
Subject category
S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS; S47: OTHER INSTRUMENTATION;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ALGORITHMS; ALUMINIUM; ELECTRIC POTENTIAL; ELECTRODES; IMAGE PROCESSING; IMPEDANCE; KHZ RANGE; MATHEMATICAL MODELS; MATRICES; PHANTOMS; TOMOGRAPHY
Descriptors DEC
DIAGNOSTIC TECHNIQUES; ELEMENTS; FREQUENCY RANGE; MATHEMATICAL LOGIC; METALS; MOCKUP; PROCESSING; STRUCTURAL MODELS