Published July 1, 1993 | Version v1
Journal article

Calibration of electron spectrometers by using 99mTc internal conversion electron sources

  • 1. Institute of Nuclear Research, Debrecen, (Hungary)

Description

Methods of absolute resolution and efficiency calibration based on internal conversion spectroscopic measurements of 99mTc sources are described. Possibilities for determining x-ray widths are also discussed. The application of the procedures recommended here makes it possible to regain the electron spectra 'as emitted' from the sample, free from distortions due to the instruments

Additional details

Publishing Information

Journal Title
Journal of Electron Spectroscopy and Related Phenomena
Journal Volume
63
Journal Issue
1
Journal Page Range
p. 31-42.
ISSN
0368-2048
CODEN
JESRAW