Feasibility of in vivo thyroid 131I monitoring with an imaging plate
Description
A new in vivo thyroid 131I monitoring method was devised by using an imaging plate (IP). A thyroid image obtained with a realistic neck-thyroid phantom showed a unique shape characteristic of the thyroid gland. A 131I thyroid imaging allows visual confirmation of thyroid accumulation of 131I. The detection limit of the IP system of 290 Bq was about ((1)/(100)) of the screening level of 30 kBq in cases of public emergencies. The IP system is applicable for thyroid 131I monitoring
Additional details
Identifiers
- PII
- S0969804301000963;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 55
- Journal Issue
- 4
- Journal Page Range
- p. 513-516
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- Canada
- INIS RN
- 33027056
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ANTITHYROID DRUGS; BIOMEDICAL RADIOGRAPHY; IN VIVO; IODINE 131; IONOGRAPHIC IMAGING; MEDICAL SURVEILLANCE; MONITORING; RADIOTHERAPY; THYROID
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BIOMEDICAL RADIOGRAPHY; BODY; DAYS LIVING RADIOISOTOPES; DIAGNOSTIC TECHNIQUES; DRUGS; ENDOCRINE GLANDS; GLANDS; INTERMEDIATE MASS NUCLEI; IODINE ISOTOPES; ISOTOPES; MEDICINE; NUCLEAR MEDICINE; NUCLEI; ODD-EVEN NUCLEI; ORGANS; RADIOISOTOPES; RADIOLOGY; THERAPY
Optional Information
- Copyright
- Copyright (c) 2001 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.