A high resolution electrostatic time-of-flight spectrometer with adiabatic magnetic collimation
Description
A new type of spectrometer for low energy charged particles is presented. It consists of an adiabatic magnetic collimation and two filters: an electrostatic retarding potential to set a lower limit (high pass) and a time-of-flight analysis to reject high energy charged particles (low pass). Both filters are only limited in their resolution by the efficiency of the adiabatic magnetic collimation. The proof of this principle is demonstrated by a pilot measurement on the K conversion line of 83mKr. Possible applications to pulsed and continuous electron sources are discussed with the emphasis on the investigation of the β spectrum of T2 to deduce information on the mass of the electron antineutrino and possible anomalies in the β spectrum. In this context design parameters of a spectrometer with a resolving power of E/ΔE=20 000 and a luminosity of A ΔΩ/4π=4 cm2 for 20 keV electrons are given
Additional details
Identifiers
- PII
- S0168900298012637;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 421
- Journal Issue
- 1-2
- Journal Page Range
- p. 256-265
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34035386
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COLLIMATORS; DESIGN; ELECTRON SPECTROMETERS; KEV RANGE 10-100; MAGNETIC FIELD CONFIGURATIONS; PERFORMANCE TESTING; TIME-OF-FLIGHT SPECTROMETERS
- Descriptors DEC
- ENERGY RANGE; KEV RANGE; MEASURING INSTRUMENTS; SPECTROMETERS; TESTING