Published February 1990 | Version v1
Journal article

A CT-based computerized treatment planning system for I-125 stereotactic brain implants

Description

A computer program has been developed at the University of California, San Francisco, as an aid in planning and evaluating stereotactic brain implants made with 125I seeds. The program allows images of seeds and catheters to be positioned in the target volume revealed by CT. It then generates and displays the resulting isodose distributions. Catheters may be changed interactively until an optimum implant is achieved. From the geometry of a stereotactic implant frame as measured by CT, the program calculates the approach angles of the catheters in the frame coordinate system. After the seeds are implanted, films made with a fiducial marker box can be used to generate true seed positions and hence true isodoses. This paper describes mathematically the geometrical transformations used by the program, and also outlines its many features and options. In its first 2 years of use the program has proved to be a valuable contributor to improved patient care

Additional details

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
18
Journal Issue
2
Series
Int. J. Radiat. Oncol., Biol. Phys.
Journal Page Range
445-454
ISSN
0360-3016
CODEN
IOBPD