Performance of FBK SiPMs coupled to PETA3 read-out ASIC for PET application
Creators
- 1. Fondazione Bruno Kessler (FBK), Trento (Italy)
- 2. University of Heidelberg, Heidelberg (Germany)
- 3. RWTH/UKA-University, Aachen (Germany)
- 4. Philips Research Laboratories, Aachen (Germany)
Description
In this paper we show the energy and timing resolution performances of FBK SiPMs coupled to the PETA3 ASIC for PET application. We developed a measurement set-up to characterize single SiPMs coupled to scintillator exploiting the detector stack developed within the HYPERImage project. In this way we are able to characterize the combined SiPM/ASIC performance with the same signal chain (from the sensor to the ASIC board) used in the PET system. We show that using two scintillator detectors, composed of a 3×3×5mm3 LYSO crystal coupled to a 3×3mm2 SiPM, an intra-stack CRT of about 200 ps FWHM can be obtained
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2012.10.012Additional details
Identifiers
- DOI
- 10.1016/j.nima.2012.10.012;
- PII
- S0168-9002(12)01144-8;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 718
- Journal Page Range
- p. 345-346
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 12. Pisa meeting on advanced detectors
- Dates
- 20-26 May 2012
- Place
- La Biodola, Elba (Italy)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 45047899
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CRYSTALS; INTEGRATED CIRCUITS; LUTETIUM COMPOUNDS; PERFORMANCE; POSITRON COMPUTED TOMOGRAPHY; READOUT SYSTEMS; RESOLUTION; SENSORS; SI SEMICONDUCTOR DETECTORS; SIGNALS; SOLID SCINTILLATION DETECTORS; YTTRIUM SILICATES
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; ELECTRONIC CIRCUITS; EMISSION COMPUTED TOMOGRAPHY; MEASURING INSTRUMENTS; MICROELECTRONIC CIRCUITS; OXYGEN COMPOUNDS; RADIATION DETECTORS; RARE EARTH COMPOUNDS; SCINTILLATION COUNTERS; SEMICONDUCTOR DETECTORS; SILICATES; SILICON COMPOUNDS; TOMOGRAPHY; TRANSITION ELEMENT COMPOUNDS; YTTRIUM COMPOUNDS
Optional Information
- Copyright
- Copyright (c) 2012 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.