Published July 21, 2002
| Version v1
Journal article
On the efficiency calibration of a drum waste assay system
Description
The efficiency calibration of a gamma spectroscopy waste assay system, constructed by IFIN-HH, was performed. The calibration technique was based on the assumption of a uniform distribution of the source activity in the drum and also a uniform sample matrix. A collimated detector (HPGe--20% relative efficiency) placed at 30 cm from the drum was used. The detection limit for 137Cs and 60Co is approximately 45 Bq/kg for a sample of about 400 kg and a counting time of 10 min. A total measurement uncertainty of -70% to +40% was estimated
Additional details
Identifiers
- PII
- S0168900201021982;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 487
- Journal Issue
- 3
- Journal Page Range
- p. 661-666
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34008255
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; CESIUM 137; COBALT 60; GAMMA SPECTROSCOPY; HIGH-PURITY GE DETECTORS; RADIATION DOSE DISTRIBUTIONS; RADIOACTIVE WASTES; SENSITIVITY
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; COBALT ISOTOPES; GE SEMICONDUCTOR DETECTORS; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MATERIALS; MEASURING INSTRUMENTS; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; RADIATION DETECTORS; RADIOACTIVE MATERIALS; RADIOISOTOPES; SEMICONDUCTOR DETECTORS; SPECTROSCOPY; WASTES; YEARS LIVING RADIOISOTOPES
Optional Information