Optimization of Direct Current-Enhanced Radiofrequency Ablation: An Ex Vivo Study

  • 1. RWTH Aachen University, Applied Medical Engineering, Helmholtz-Institute Aachen (Germany)
  • 2. Nara Medical University, Department of Radiology (Japan)

Description

The purpose of this study was to investigate the optimal setting for radiofrequency (RF) ablation combined with direct electrical current (DC) ablation in ex vivo bovine liver. An electrical circuit combining a commercially available RF ablation system with DC was developed. The negative electrode of a rectifier that provides DC was connected to a 3-cm multitined expandable RF probe. A 100-mH inductor was used to prevent electrical leakage from the RF generator. DC was applied for 15 min and followed by RF ablation in freshly excised bovine livers. Electric current was measured by an ammeter. Coagulation volume, ablation duration, and mean amperage were assessed for various DC voltages (no DC, 2.2, 4.5, and 9.0 V) and different RF ablation protocols (stepwise increase from 40 to 80 W, 40 W fixed, and 80 W fixed). Results were compared using Kruskal-Wallis and Mann-Whitney U test. Applying DC with 4.5 or 9.0 V, in combination with 40 W fixed or a stepwise increase of RF energy, resulted in significantly increased zone of ablation size compared with 2.2 V or no DC (P = 0.009). At 4.5 V DC, the stepwise increase of RF energy resulted in the same necrosis size as a 40 W fixed protocol (26.6 ± 3.9 vs. 26.5 ± 4.0 ml), but ablation duration was significantly decreased (296 ± 85 s vs. 423 ± 104 s; P = 0.028). Mean amperage was significantly lower at 4.5 V compared with 9.0 V (P = 0.028). Combining a stepwise increase of RF energy with a DC voltage of 4.5 V is most appropriate to increase coagulation volume and to minimize procedure time.

Additional details

Identifiers

Publishing Information

Journal Title
Cardiovascular and Interventional Radiology
Journal Volume
33
Journal Issue
5
Journal Page Range
p. 1028-1032
ISSN
0174-1551
CODEN
CAIRDG

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
42028980
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ABLATION; CATTLE; DIRECT CURRENT; LIVER; OPTIMIZATION; RADIOWAVE RADIATION
Descriptors DEC
ANIMALS; BODY; CURRENTS; DIGESTIVE SYSTEM; DOMESTIC ANIMALS; ELECTRIC CURRENTS; ELECTROMAGNETIC RADIATION; GLANDS; MAMMALS; ORGANS; RADIATIONS; RUMINANTS; VERTEBRATES

Optional Information

Copyright
Copyright (c) 2010 Springer Science+Business Media, LLC and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE)