Published November 1985
| Version v1
Report
Restricted
An on-line actinide-in-air monitor to operate at concentrations below 0.1 ICRP MPCsub(a)
Description
A prototype on-line actinide-in-air monitoring system has been built which has sufficient sensitivity to determine average concentrations as low as 0.04 dpm m-3 (8 x 10-3 ICRP MPCsub(a)) over an 8 h shift. Routine measurements at 0.3 dmp m-3 (0.075 ICRP MPCsub(a)) can be made with a probability of < 1% per shift of a false alarm, even in the presence of unusually large concentrations of radon daughter products. The system uses a combination of inertial particle size separation, alpha energy discrimination and isotope decay analysis to achieve this performance, which is between 10 and 100 times better than commercially available equipment. (U.K.)
Availability note (English)
MF available from INIS under the Report Number.
Files
Additional details
Publishing Information
- Imprint Pagination
- 41 p.
- Report number
- CEGB/TPRD/B--0731/N85
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 17065715
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACTINIDES; AIR SAMPLERS; ALPHA DETECTION; RADIATION MONITORING; RADIATION MONITORS; RADIOMETRIC ANALYSIS
- Descriptors DEC
- CHARGED PARTICLE DETECTION; CHEMICAL ANALYSIS; DETECTION; ELEMENTS; EQUIPMENT; LABORATORY EQUIPMENT; MEASURING INSTRUMENTS; METALS; MONITORING; MONITORS; QUANTITATIVE CHEMICAL ANALYSIS; RADIATION DETECTION; SAMPLERS