Improved method for the stabilization of NaI-photomultiplier gamma detectors against thermal and other drift

Description

Alpha peaks have been used as part of servo systems to stabilize NaI-photomultiplier gamma detectors against drift. However, alpha peaks shift with temperature change differently than do gamma peaks, thus spoiling what would otherwise be a workable scheme for stabilizing against probably the most serious source of NaI-p.m. detector drift, namely thermal effects. It has been found possible to accurately compensate for the difference in the shift with temperature versus gamma peaks using the signal derived from a thermistor in thermal contact with the NaI crystal to control the bias of a discriminator in the servo circuit. The servo circuit utilizing this principle has been used in commercial multichannel analyzers of the type intended for field use under adverse ambient conditions

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01 as DE85011830.

Files

17003629.pdf

Files (260.7 kB)

Name Size Download all
md5:bd8d04e7e22164277027c68f3705e849
260.7 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
19 p.
Report number
BNL--36395

Conference

Title
6. symposium on X- and gamma-ray sources and applications.
Dates
21-23 May 1985.
Place
Ann Arbor, MI (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
17003629
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
GAMMA DETECTION; NAI DETECTORS; PERFORMANCE; PHOTOMULTIPLIERS; STABILITY; TEMPERATURE DEPENDENCE
Descriptors DEC
DETECTION; MEASURING INSTRUMENTS; PHOTOTUBES; RADIATION DETECTION; RADIATION DETECTORS; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS

Optional Information

Secondary number(s)
CONF-850533--7.