Published July 2008 | Version v1
Journal article

Obtainment of precise experimental data on diffractometers with 2D and point detectors

Creators

  • 1. Russian Academy of Sciences, Shubnikov Institute of Crystallography (Russian Federation)

Description

A measurement method based on the use of diffractometers with a point detector is proposed that makes it possible to reduce the measurement time by a factor of about 2 in comparison with the conventional experiment. Measures are proposed that allow one to increase the measurement precision for diffractometers with a 2D CCD detector. Refinement of the model of independent atoms from the experimental data obtained by these methods is characterized by consistency between the experimental and calculated structure factors at a level of 1-1.5%. A further significant decrease in refinement Rfactors in going to the multipole model indicates that the data precision is sufficient for studying the charge density in crystals.

Additional details

Identifiers

Publishing Information

Journal Title
Crystallography Reports
Journal Volume
53
Journal Issue
4
Journal Page Range
p. 717-720
ISSN
1063-7745
CODEN
CYSTE3

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
44010941
Subject category
S75: CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY;
Descriptors DEI
ACCURACY; CHARGE DENSITY; COMPARATIVE EVALUATIONS; CRYSTALS; DIFFRACTOMETERS; STRUCTURE FACTORS
Descriptors DEC
DIMENSIONLESS NUMBERS; EVALUATION; MEASURING INSTRUMENTS

Optional Information

Copyright
Copyright (c) 2008 Pleiades Publishing, Ltd.