Published July 1991 | Version v1
Journal article

Dose rate in irregularly shaped high energy photon beams

  • 1. M. Sklodowska-Curie Memorial Inst., Gliwice (Poland). Center of Oncology
  • 2. M. Sklodowska-Curie Memorial Hospital, Poznan (Poland). Center of Oncology

Description

A quick method of calculating the dose-rate distribution in irregularly shaped beams is presented. This method is partly based on the Cunningham model. The proposed modifications make it possible to avoid the scatter functions in favor of direct use of tissue-phantom ratios and field size coefficients. The problem has been evaluated for photons generated by 10 and 23 MV linacs. The calculated dose rates have been compared with measured ones, and good accuracy has been achieved. (orig.)

Abstract (German)

Eine Methode wird vorgestellt, die eine schnelle Berechnung der Dosisleistungsverteilung bei unregelmaessigen Strahlen erlaubt. Diese Methode ist teilweise an das Cunningham-Modell angelehnt. Durch die vorgeschlagene Modifizierung ist es moeglich, anstelle der Streufunktionen direkt mit dem Verhaeltnis Gewebe-Phantom und den Feldgroessenkoeffizienten zu arbeiten. Die Werte wurden fuer die Photonenstrahlung der Geraete Linac 10 und 23 MV bestimmt. Die errechneten und die gemessenen Dosisleistungen wurden verglichen, wobei sich eine gute Uebereinstimmung ergab. (orig.)

Additional details

Publishing Information

Journal Title
Strahlentherapie und Onkologie
Journal Volume
167
Journal Issue
7
Series
Strahlenther. Onkol.
Journal Page Range
416-418
ISSN
0179-7158
CODEN
STONE