Published February 2001 | Version v1
Journal article

Real-time monitoring and verification of in vivo high dose rate brachytherapy using a pinhole camera

  • 1. Department of Radiation Oncology, University of Alabama at Birmingham, 619 South 19th Street, Birmingham, Alabama 35233 (United States)

Description

We investigated a pinhole imaging system for independent in vivo monitoring and verification of high dose rate (HDR) brachytherapy treatment. The system consists of a high-resolution pinhole collimator, an x-ray fluoroscope, and a standard radiographic screen-film combination. Autofluoroscopy provides real-time images of the in vivo Ir-192 HDR source for monitoring the source location and movement, whereas autoradiography generates a permanent record of source positions on film. Dual-pinhole autoradiographs render stereo-shifted source images that can be used to reconstruct the source dwell positions in three dimensions. The dynamic range and spatial resolution of the system were studied with a polystyrene phantom using a range of source strengths and dwell times. For the range of source activity used in HDR brachytherapy, a 0.5 mm diameter pinhole produced sharp fluoroscopic images of the source within the dynamic range of the fluoroscope. With a source-to-film distance of 35 cm and a 400 speed screen-film combination, the same pinhole yielded well recognizable images of a 281.2 GBq (7.60 Ci) Ir-192 source for dwell times in the typical clinical range of 2 to 400 s. This 0.5 mm diameter pinhole could clearly resolve source positions separated by lateral displacements as small as 1 mm. Using a simple reconstruction algorithm, dwell positions in a phantom were derived from stereo-shifted dual-pinhole images and compared to the known positions. The agreement was better than 1 mm. A preliminary study of a patient undergoing HDR treatment for cervical cancer suggests that the imaging method is clinically feasible. Based on these studies we believe that the pinhole imaging method is capable of providing independent and reliable real-time monitoring and verification for HDR brachytherapy

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
28
Journal Issue
2
Journal Page Range
p. 167-173
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
35001816
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
DOSE RATEMETERS; DOSIMETRY; IMAGE PROCESSING; NEOPLASMS; RADIATION MONITORING; RADIATION SOURCE IMPLANTS; RADIOTHERAPY
Descriptors DEC
DISEASES; IMPLANTS; MEDICINE; MONITORING; NUCLEAR MEDICINE; PROCESSING; RADIATION SOURCES; RADIOLOGY; THERAPY

Optional Information

Notes
(c) 2001 American Association of Physicists in Medicine.