Published February 1993 | Version v1
Journal article

A new uranium automatic analyzer

  • 1. Beijing Research Inst. of Chemical Engineering and Metallurgy, CNNC (China)

Description

A new uranium automatic analyzer based on the flow injection analysis (FIA) principle has been developed. It consists of a multichannel peristaltic pump, an injection valve, a photometric detector, a single-chip microprocessor system and electronic circuit. The new designed multifunctional auto-injection valve can automatically change the injection volume of the sample and the channels so that the determination ranges and items can easily be changed. It also can make the instrument vary the FIA operation modes that it has functions of a universal instrument. A chromatographic column with extractant-containing resin was installed in the manifold of the analyzer for the concentration and separation of trace uranium. The 2-(5-bromo-2-pyridylazo)-5-diethyl-aminophenol (Br-PADAP) was used as colour reagent. Uranium was determined in the aqueous solution by adding cetyl-pyridium bromide (CPB). The uranium in the solution in the range 0.02-500 mg · L-1 can be directly determined without any pretreatment. A sample throughput rate of 30-90 h-1 and reproducibility of 1-2% were obtained. The analyzer has been satisfactorily applied to the laboratory and the plant

Additional details

Publishing Information

Journal Title
Uranium Mining and Metallurgy
Journal Volume
12
Journal Issue
1
Journal Page Range
p. 38-45, 50.
ISSN
1000-8063
CODEN
YOUKEM