Published 1995 | Version v1
Journal article

Patient dose and image quality for computed tomography in several Dutch hospitals

  • 1. University Hospital, Leiden (Netherlands). Depts. of Clinical Oncology and Diagnostic Radiology
  • 2. OLVG Hospital, Amsterdam (Netherlands)
  • 3. TNO, Rijswijk (Netherlands). Medical Biological Lab.

Description

In several Dutch hospitals a survey of computed tomography (CT) techniques was performed at 14 CT scanners. In this survey the computed tomography dose index (CTDI) was measured free-in-air with a pencil type ionisation chamber. These measurements yield dose values related to a single CT scan of one slice whilst radiation exposure of patients due to a complete CT examination should preferably be expressed in terms of organ and tissue doses or effective dose. Organ and effective doses were derived by application of published organ dose conversion factors. Effective doses varied from 0.7 mSv to 2.2 mSv for head scans, from 6.0 mSv to 13 mSv for chest scans and from 7.1 mSv to 21 mSv for standard abdomen scans. The survey also included the assessment of parameters related to image quality i.e. high contrast resolution and noise (standard deviation, ω, of Hounsfield units) associated with the CT scanners and protocols at the different locations. Image quality in terms of noise or resolution and effective dose did not reveal a significant correlation for the different CT scanners. (Author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
57
Journal Issue
1-4
Journal Page Range
p. 129-133.
ISSN
0144-8420
CODEN
RPDODE

Conference

Title
Quality control and radiation protection of the patient in diagnostic radiology and nuclear medicine conference.
Dates
29 Sep - 1 Oct 1993.
Place
Grado (Italy).

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
28000988
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ABDOMEN; CHEST; COMPUTERIZED TOMOGRAPHY; HEAD; HOSPITALS; IMAGES; NETHERLANDS; NOISE; PATIENTS; PHANTOMS; RADIATION DOSES; SPATIAL RESOLUTION
Descriptors DEC
ANIMALS; BODY; BODY AREAS; BUILDINGS; DEVELOPED COUNTRIES; EUROPE; MAMMALS; MAN; MEDICAL ESTABLISHMENTS; MOCKUP; PRIMATES; RESOLUTION; STRUCTURAL MODELS; TOMOGRAPHY; VERTEBRATES; WESTERN EUROPE

Optional Information

Secondary number(s)
EUR--15257.