Published February 1986 | Version v1
Report Open

Long-term changes in the sensitivity of quadrupole mass spectrometers

Description

We routinely use quadrupole mass spectrometers (QMS) to monitor vacuum conditions, gas purity, and plasma-wall interactions in the Tokamak Fusion Test Reactor (TFTR) at Princeton. Two QMS systems have been operating on TFTR continuously for a two-year period. Both QMS systems are absolutely calibrated at weekly intervals using a six-part standard gas mixture. The calibration procedure is based on the use of transfer standards (ion gauge and capacitance manometer) that are calibrated against a primary standard (spinning rotor gauge) on an external vacuum system. We have identified variations in the efficiency of the QMS ionizer and drifts in the sensitivity of the electron multiplier ion detector to be the major reasons for the observed changes in overall OMS sensitivity. Weekly variations in sensitivity greater than 100% have been observed following system bakeout at 1500C and with the use of rhenium filaments which were initially in the QMS ionizer. Operation of the QMS systems with tungsten filaments and at constant temperature has yielded more stable operation with weekly sensitivity changes generally being less than 10%. 7 refs., 7 figs

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01; 1 as DE86007531.

Files

17063721.pdf

Files (366.8 kB)

Name Size Download all
md5:72e9589035743e3474c14b421f0cf5f8
366.8 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
20 p.
Report number
PPPL--2305

Conference

Title
32. national vacuum symposium and topical conference.
Dates
19-22 Nov 1985.
Place
Houston, TX (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
17063721
Subject category
S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CALIBRATION; IMPURITIES; MASS SPECTROMETERS; PLASMA DIAGNOSTICS; SENSITIVITY; TFTR TOKAMAK
Descriptors DEC
CLOSED PLASMA DEVICES; MEASURING INSTRUMENTS; SPECTROMETERS; THERMONUCLEAR DEVICES; TOKAMAK DEVICES

Optional Information

Notes
Portions of this document are illegible in microfiche products. Original copy available until stock is exhausted.
Secondary number(s)
CONF-851174--40.