Published June 2000
| Version v1
Journal article
A study on the reduction of exposure dose and contrast improvement by use of heavy elements filter
Description
This work was on the reduction of exposure dose and contrast improvement by use of heavy elements filter. From the result of experimental evaluation, it was found that the beam harding of X-ray was not showed in Ho and Gd, heavy elements filters, contrast to Cu and Al filters in which the harding showed. And the ratio of transit dose to surface dose and the load of X-ray tube increase in order of Al, Cu, Gd and Ho, respectively. The contrast of X-ray images using the intensifying screen and the input phosphor showed the higher value in order of Cu, Al, Gd and Ho. Therefore, in the case of using contrast media and phosphor in region of diagnostic radiology, X-ray image quality depends primarily on kVp and heavy elements filters.=20
Additional details
Publishing Information
- Journal Title
- Journal of the Korean Society of Radiological Technology
- Journal Volume
- 23
- Journal Issue
- 1
- Series
- 15 refs, 14 figs
- Journal Page Range
- p. 91-96
- ISSN
- 1226-2854
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 31060036
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DIAGNOSTIC TECHNIQUES; FILTERS; GADOLINIUM; HOLMIUM; PATIENTS; RADIATION DOSES; RADIATION PROTECTION; REDUCTION
- Descriptors DEC
- CHEMICAL REACTIONS; DIAGNOSTIC TECHNIQUES; DOSES; ELEMENTS; MEDICINE; METALS; NUCLEAR MEDICINE; RADIOLOGY; RARE EARTHS