Mass spectrometer of the VEGA-1 automatic space station

  • 1. AN SSSR, Moscow (USSR). Inst. Geokhimii i Analiticheskoj Khimii

Description

Mass-spectrometric installation designed to determine chemical composition of Venus cloud veil set on the launching vehicle of VEGA-1 automatic space station is described. Spectrometric installation consists of a sampler, air filter system, collector-pyrolyzer and three-dimensional trap type hyperboloid mass-spectrometer. Great attention was paid to reduction of background signal, it has allowed to obtain ∼1016 dynamic range at 5x10-12 Torr threshold sensitivity

Additional details

Additional titles

Original title (Russian)
Масс-спектрометр автоматической межпланетной станции ''Вега-1ь

Publishing Information

Journal Title
Pribory i Tekhnika Ehksperimenta
Journal Issue
no.4
Series
Prib. Tekh. Ehksp.
ISSN
0032-8162
CODEN
PRTEA