Published November 2017 | Version v1
Journal article

Designing a novel phantom for daily QA of cyberknife M6 robotic radiosurgery system

  • 1. Department of Medical Physics and Radiation Safety, Amrita Institute of Medical Science and Research Center, Kochi (India)
  • 2. Department of Physics, Karunya University, Coimbatore (India)

Description

Performing daily QA on Cyberknife system includes routine linac output constancy and constancy check for energy. Placing the ionization chamber at 80 cm SAD as specified by the equipment supplier for Cyberknife system is difficult since the system is non isocentric. However, birdcage assembly, an assembly and detector mount for holding the ionization chamber at a fixed distance of 80 CM from the the radiation source can be used to monitor the output. But the measurement chamber is placed in air which is not suitable for 6MV photon beams. Hence a new phantom has been designed which can be fixed to birdcage assembly to monitor output and energy constancy in the medium (Polystyrene)

Additional details

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
42
Journal Issue
suppl.1
Journal Page Range
p. 154-155
CODEN
JMPHFE

Conference

Title
17. Asia-Oceania Congress of Medical Physics; 38. annual conference of Association of Medical Physicists of India
Acronym
AOCMP-AMPICON 2017
Dates
4-7 Nov 2017
Place
Jaipur (India)

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
49051779
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
QUALITY ASSURANCE; RADIATION DOSES; RADIOTHERAPY; ROBOTS; SURGERY
Descriptors DEC
DOSES; EQUIPMENT; MANAGEMENT; MEDICINE; NUCLEAR MEDICINE; QUALITY MANAGEMENT; RADIOLOGY; THERAPY