Description
Performing daily QA on Cyberknife system includes routine linac output constancy and constancy check for energy. Placing the ionization chamber at 80 cm SAD as specified by the equipment supplier for Cyberknife system is difficult since the system is non isocentric. However, birdcage assembly, an assembly and detector mount for holding the ionization chamber at a fixed distance of 80 CM from the the radiation source can be used to monitor the output. But the measurement chamber is placed in air which is not suitable for 6MV photon beams. Hence a new phantom has been designed which can be fixed to birdcage assembly to monitor output and energy constancy in the medium (Polystyrene)
Additional details
Publishing Information
- Journal Title
- Journal of Medical Physics
- Journal Volume
- 42
- Journal Issue
- suppl.1
- Journal Page Range
- p. 154-155
- CODEN
- JMPHFE
Conference
- Title
- 17. Asia-Oceania Congress of Medical Physics; 38. annual conference of Association of Medical Physicists of India
- Acronym
- AOCMP-AMPICON 2017
- Dates
- 4-7 Nov 2017
- Place
- Jaipur (India)
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 49051779
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- QUALITY ASSURANCE; RADIATION DOSES; RADIOTHERAPY; ROBOTS; SURGERY
- Descriptors DEC
- DOSES; EQUIPMENT; MANAGEMENT; MEDICINE; NUCLEAR MEDICINE; QUALITY MANAGEMENT; RADIOLOGY; THERAPY