Demonstration of Johnson noise thermometry with all-superconducting quantum voltage noise source

  • 1. National Institute of Advanced Industrial Science and Technology, 1-1-1 Umezono, Tsukuba, Ibaraki 305-8568 (Japan)

Description

We present a Johnson noise thermometry (JNT) system based on an integrated quantum voltage noise source (IQVNS) that has been fully implemented using superconducting circuit technology. To enable precise measurement of Boltzmann's constant, an IQVNS chip was designed to produce intrinsically calculable pseudo-white noise to calibrate the JNT system. On-chip real-time generation of pseudo-random codes via simple circuits produced pseudo-voltage noise with a harmonic tone interval of less than 1 Hz, which was one order of magnitude finer than the harmonic tone interval of conventional quantum voltage noise sources. We estimated a value for Boltzmann's constant experimentally by performing JNT measurements at the temperature of the triple point of water using the IQVNS chip