Published January 1998 | Version v1
Journal article

Resistive anodes for one-dimensional position-sensitive detector

  • 1. Tsinghua Univ., Beijing (China). Dept. of Applied Physics

Description

The theory, the performance and the application of resistive anodes for one-dimensional position-sensitive detector are talked about. The one-dimensional resistive anode with the position resolution better than 200 μm and the linear coefficient between the energy and position γ 0.99962138 have been achieved. They work very well in the instrument of Electron Momentum Spectroscopy (EMS)

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
18
Journal Issue
1
Journal Page Range
p. 54-56.
ISSN
0258-0934
CODEN
HDYUEC