Published June 2014 | Version v1
Journal article

SU-E-T-149: Electron Beam Profile Differences Between Elekta MLCi2 and Elekta Agility Treatment Heads

  • 1. Sutter Medical Foundation, Roseville, CA 95661 (United States)
  • 2. Sutter Shared Services S3, Sacramento, CA (United States)

Description

Purpose: To report and investigate observed differences in electron beam profiles at various energies/applicators between Elekta MLCi2 and Agility treatment head on Elekta Infinity LINAC Methods: When we upgraded from MLCi2 to Agility on one of our Elekta Infinity LINAC's, electron beam PDDs and profiles were acquired for comparison purpose. All clinical electron energies (6/9/12/15/12/18 MeV) and electron applicators (6/10/14/20/25 square) were included in measurement. PDDs were acquired at 100 SSD in water (PTW MP3 water tank) with a plane-parallel ion chamber (PTW Roos). X and Y Profiles were acquired using IC Profiler (Sun Nuclear Corp.) at 1cm and maximum PDD depths (water equivalent). Results: All PDD curves match very well between MLCi2 and Agility treatment head. However, some significant differences on electron profiles were found. On Agility, even after increasing the default auto-tracking offset values for backup diaphragms in Y and MLC in X by 2.8 cm (the maximum allowed change is 3.0 cm), electron profiles still have rounder shoulders comparing to corresponding MLCi2 profiles. This difference is significantly more pronounced at larger applicators (20 and 25 square), for all electron energies. Conclusion: The significant design change between MLCi2 and Agility beam limiting device seems to affect exit electron beam profiles. In IEC1217 X direction, the main change on Agility is the removal of the original MLCi2 X backup diaphragms and replacing it with MLC leaves; In Y direction, the main change is the radius and materials on Y backup diaphragms

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
41
Journal Issue
6
Journal Page Range
p. 257
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46046218
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Descriptors DEI
DIAPHRAGM; ELECTRON BEAMS; HEAD; IONIZATION CHAMBERS; LINEAR ACCELERATORS; RADIOTHERAPY
Descriptors DEC
ACCELERATORS; BEAMS; BODY; LEPTON BEAMS; MEASURING INSTRUMENTS; MEDICINE; MUSCLES; NUCLEAR MEDICINE; ORGANS; PARTICLE BEAMS; RADIATION DETECTORS; RADIOLOGY; THERAPY

Optional Information

Notes
(c) 2014 American Association of Physicists in Medicine