Hybrid monitor for both beam position and tilt of pulsed high-current beams
Description
A Hybrid beam monitor, integrated with both azimuthal and axial B-dot probes, was designed for simultaneous measurement of both beam position and beam angle for pulsed high-current beams at the same location in beam pipe. The output signals of axial B-dot probes were found to be mixed with signals caused by transverse position deviation. In order to eliminate the unwanted signals, an elimination method was developed and its feasibility tested on a 50-Ω coaxial line test stand. By this method, a waveform, shape-like to that of input current and proportional to the tilt angle, was simulated and processed by following integration step to achieve the tilt angle. The tests showed that the measurement error of displacement and tilt angle less than 0.3 mm and 1.5 mrad, respectively. The latter error could be reduced with improved probes by reducing the inductance of the axial B-dot probe, but the improvement reached a limit due to some unknown systemic mechanism
Additional details
Identifiers
- DOI
- 10.1063/1.4895643;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 85
- Journal Issue
- 9
- Journal Page Range
- p. 093305-093305.6
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46020644
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BEAM CURRENTS; BEAM MONITORS; BEAM POSITION; ERRORS; PROBES; PULSES; SIGNALS; WAVE FORMS
- Descriptors DEC
- CURRENTS; MEASURING INSTRUMENTS; MONITORS
Optional Information
- Notes
- (c) 2014 AIP Publishing LLC