Published 2008 | Version v1
Miscellaneous Open

Evaluating the accuracy of uranium isotope amount ratio measurements performed by a quadrupole and a multi-collector magnetic sector inductively coupled plasma mass spectrometers for nuclear safeguards

  • 1. Instituto de Pesquisas Energeticas e Nucleares (IPEN), Av. Lineu Prestes 2242, Butanta, 05508-000 Sao Paulo, SP (Brazil)
  • 2. Institute for Reference Materials and Measurements (IRMM), Joint Research Center, European Commission, Retieseweg 111, 2440 Geel (Belgium)

Description

The n(U235)/n(U238) isotope amount ratio in a set of samples was measured using two modern analytical techniques: quadrupole inductively coupled plasma mass spectrometry (ICP-QMS) and multi-collector magnetic sector inductively coupled plasma mass spectrometry (MC-ICPMS). The measured ratios were compared to the certified ratios provided by the high accuracy gas source mass spectrometry (GSMS). The components of the uncertainty were identified and their contribution to the combined standard uncertainty was estimated using the recommendations of the ISO-GUM guide. The values of the measurement uncertainty and bias were determined and then compared to the International Target Values for Measurement Uncertainties in Safeguarding Nuclear Materials. It appears that only the measurements performed by MC-ICPMS can meet the stringent requirements of international nuclear safeguards. (authors)

Availability note (English)

Available from INIS in electronic form

Files

40032678.pdf

Files (55.5 kB)

Name Size Download all
md5:30d0e3ca5a031cc9400c45f92ffcfbba
55.5 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
4 p.
Report number
INIS-FR--09-0057

Conference

Title
Nuclear fuel cycle for a sustainable future
Acronym
Atalante 2008
Dates
19-23 May 2008
Place
Montpellier (France)

Optional Information

Notes
6 refs.