Technique for calibrating V probes

Description

V probes designed by Maxwell Laboratories, Inc. (MLI) have been extensively used at the Air Force Weapons Laboratory to measure the voltage of fast capacitor banks. The straightforward calibration method of pulsing a high voltage across the voltage probe typically produces only millivolts of output signal. Based on a suggestion by MLI, a new method to calibrate the probes has been devised. This technique, which uses a capacitive discharge and measures the dV/dt directly, has the advantage of producing an output signal of nearly a volt. A circuit has been designed and constructed to provide an appropriate ringing discharge which can be used to calibrate V probes. The theory of V probes and their calibration is discussed, and the construction and operation of the discharge circuit is presented. Finally, as an example of the calibration technique, details of how to calibrate a V probe are presented. A precision of 4% is readily attainable in the V probe calibration. Such accuracy should make possible routine 5% voltage measurements on pulsed power systems

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS., PC A02/MF A01.

Files

12575670.pdf

Files (182.3 kB)

Name Size Download all
md5:a1bb831da7e3e1cc0f06039fca49ce50
182.3 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
21 p.
Report number
SAND--80-1071

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
12575670
Subject category
S42: ENGINEERING;
Descriptors DEI
CALIBRATION; CAPACITORS; ELECTRIC POTENTIAL; ELECTRIC PROBES; MEASURING INSTRUMENTS; POWER TRANSMISSION LINES; TRANSIENTS
Descriptors DEC
ELECTRICAL EQUIPMENT; ENERGY STORAGE SYSTEMS; PROBES